| Author | |
|---|---|
| Abstract | :  Skin cancer treatment involving (32)P source is an easy, less expensive method of treatment limited to small and superficial lesions of approximately 1 mm deep. Bhabha Atomic Research Centre (BARC) has indigenously developed (32)P nafion-based patch source (1 cm × 1 cm) for treating skin cancer. For this source, the values of dose per unit activity at different depths including dose profiles in water are calculated using the EGSnrc-based Monte Carlo code system. For an initial activity of 1 Bq distributed in 1 cm(2) surface area of the source, the calculated central axis depth dose values are 3.62 × 10(-10) GyBq(-1) and 8.41 × 10(-11) GyBq(-1)at 0.0125 and 1 mm depths in water, respectively. Hence, the treatment time calculated for delivering therapeutic dose of 30 Gy at 1 mm depth along the central axis of the source involving 37 MBq activity is about 2.7 hrs. | 
| Year of Publication | :  1969 | 
| Journal | :  Journal of medical physics | 
| Volume | :  40 | 
| Issue | :  1 | 
| Number of Pages | :  13-7 | 
| Date Published | :  1969 | 
| ISSN Number | :  0971-6203 | 
| URL | :  http://www.jmp.org.in/article.asp?issn=0971-6203;year=2015;volume=40;issue=1;spage=13;epage=17;aulast=Sahoo | 
| DOI | :  10.4103/0971-6203.152232 | 
| Short Title | :  J Med Phys | 